a. 1 MN MTS Load Frame
The 1 MN MTS closed-loop servo-hydraulic load frame provides a vast ability to apply axial loading for determining the poroelastic properties, conducting flow measurements, and strength characterization for reservoir and caprock materials. The large capacity allows us to simulate the extreme in-situ conditions or external loading for testing.

Applications:
- Uniaxial Compression Strength (UCS): With sufficient strain gauges installed to the specimen, the UCS can be measured with the Young’s modulus and Poisson’s ratio during the loading/unloading of elastic deformation. During the failure test, AE sensors can also be instrumented.


- Triaxial Compression test: With the Hooke’s cell or the advanced triaxial cell, the strength parameters can be determined with different confining pressures.


b. THM cell
The THM cell manufactured by GCTS is a 70 MPa / 150 ºC capacity triaxial cell for measurements of the poromechanical response, two-phase flow, acoustic emission (AE) activity, and failure characteristics of rock. Using the cell lifting system, the THM cell can be placed into the 220 kips MTS load frame to apply the axial load of up to 1 MN. The standard specimen size is 50 mm in diameter and 90-115 mm long. Heat shrink tubing is used to seal the specimens from the confining fluid (hydraulic oil). Two axial LVDTs and one circumferential extensometer are installed around the specimen to measure its deformation – this can also be complemented with the direct directional strain measurements with the strain gages. 13 feedthrough channels are available for temperature, deformation, strain and AE sensors. The external heating system allows the application of elevated temperatures that are monitored with internal thermocouples. The cell has two upstream and one downstream channel for injection of multiple fluids with pressures up to 70 MPa. The channels are instrumented with the pressure transducers for accurate monitoring of pore fluid pressure and quantification of the transport and undrained properties of the tested rock.

c. 3.5 MPa GDS Cell
Axisymmetric compression tests can conduct measurements of the poroelastic properties of rock in a triaxial cell with a capacity of 3.5 MPa (GDS instruments, UK). The axial load is applied with the 50 kN load frame while confining pressure and input/output pore pressures are controlled by the pressure/volume control pumps with 4.0 MPa capacity. Two pressure transducers are installed at the input and output to monitor the inner pore pressure during the undrained response. The axial/lateral strains are measured from the corresponding LVDTs.

Applications:
- Full saturation is achieved by adopting the back-pressure saturation method, while during the procedure, Skempton’s B coefficient is measured. The B coefficient values increase along with the pore pressure at constant effective mean stress, and as the B values converge, we assume that we have reached full saturation.
- After reaching saturation, the compression test is conducted in drained and undrained conditions. Deviatoric loading can be controlled such that the loading and deformation are in the elastic range and does not damage the specimen. Besides, for drained loading, it is essential to control the loading rate to be slow enough that the deviatoric loading does not induce pore pressure buildup during loading. Loading-unloading tests for drained and undrained conditions are conducted, which allow the evaluation of drained (E) and undrained (Eu) Young’s moduli.

- As the solid matrix of rock exhibits viscous creep, the time-dependent behavior can be observed by the deformation of the specimen in drained condition, or the pore pressure buildup under undrained conditions.

d. 88 kN Load Frame
Fracture tests can be conducted in a closed-loop servo hydraulic 20-kips load frame from MTS systems with the feedback signal taken as the crack mouth opening displacement (CMOD) across the notch at a gauge length of 15 mm. During the loading, the strain gauge based transducer monitors the CMOD and is programmed to increase at a rate of 0.05 μm/s. The applied load is recorded with a load cell with a capacity of 11 kN. The scaling effects of fracture toughness and fracture process zone of rock are investigated by utilizing AE sensors and the displacement which is captured by the digital image correlation (DIC) method.

Measurements: Crack-mouth opening displacement, point load displacement, accurate (few microns resolution) 2D displacement fields, 3D monitoring of damage points.

e. Hydrostatic Compression Cell
The hydrostatic compression cell provides a robust method to determine the bulk compressibility of rock. The pressure vessel is filled with hydraulic oil, where the pressure is controlled by a hydraulic pump with 70 MPa capacity. The pressure transducer measures the hydrostatic pressure, whereas the deformation is obtained locally on the rock specimen surfaces with the sets of resistive strain gages. The drained (dry) and unjacketed bulk moduli can be determined from the jacketed and unjacketed tests, respectively.

- Jacketed test: Dry specimens are covered with polyurethane and hydrostatic pressure is applied for loading and unloading in order to measure the drained (dry) bulk modulus K.

- Unjacketed test: After conducting the jacketed test, the polyurethane membrane is carefully removed with the strain gauges preserved, which allows the confining fluid to intrude into the pores of the specimen. A hydrostatic compression test is then repeated for loading and unloading. Thus, from the strain gauge responses, the unjacketed bulk modulus (Ks’ ) can be measured.